3 • Operation and Maintenance
Doc. 031842-01 5/02
3-17
3.3
Guidelines for Method Development
Follow the procedure below to develop a method for a new sample type.
1. Select a solvent(s) (see Section 3.1.1).
2. Prepare the sample (see Section 3.1.3).
3. Press the
MENU
button to display the
MAIN
screen.
4. Move the cursor to
METHOD EDITOR
and press
ENTER.
This displays the
METHOD EDITOR
screen.
Figure 3-6. Main Screen
Figure 3-7. Method Editor Screen (Additional text shown)