ASE 100 Operator’s Manual
3-22
Doc. 031842-01 5/02
Always install the cell in the cell holder. If you install the cell
directly in the oven, the cell door will not close and the run will
not start.
While holding the
extraction cell at a
slight angle, position
the bottom end cap
under the bottom
latch of the cell
holder.
Raise the cell so that
the bottom end cap
lifts the bottom latch.
Straighten the cell
until it is vertical.
Now, lower the cell
until the top end cap
rests on the top latch
of the cell holder.