ASE 100 Operator’s Manual
4-2
Doc. 031842-01 5/02
Error 002
System pressure low.
Cause:
Insufficient nitrogen gas pressure applied to the system.
Action:
Check the nitrogen supply; if it is low, replace it. Regulate the
supply to between 1.03 and 1.38 MPa (150 and 200 psi);
1.03 MPa (150 psi) is recommended.
Action:
Check the
REGULATORS
screen (see Section C.2.3). The
SYSTEM
field should read 0.34
±
0.02 MPa (50
±
3 psi); if it
does not, adjust the pressure regulator for the system gas to
0.34 MPa (50 psi).
Action:
If the error message appears again, contact Dionex for
assistance.
Error 003
Oven temperature low.
Cause:
The heater cable connection to the power supply is loose, or a
heater component malfunctioned.
Action:
Contact Dionex for assistance.
Error 004
Oven temperature high
Cause:
The heater cable connection to the power supply is loose, or a
heater component malfunctioned.
Action:
Contact Dionex for assistance.
Summary of Contents for ASE 100
Page 8: ...ASE 100 Accelerated Solvent Extractor vi Doc 031842 01 5 02...
Page 24: ...ASE 100 Operator s Manual 2 10 Doc 031842 01 5 02 Figure 2 4 ASE 100 Rear Panel...
Page 32: ...ASE 100 Operator s Manual 2 18 Doc 031842 01 5 02...
Page 42: ...ASE 100 Operator s Manual 3 10 Doc 031842 01 5 02 Figure 3 2 Inspecting Cell O Rings...
Page 60: ...ASE 100 Operator s Manual 3 28 Doc 031842 01 5 02...
Page 69: ...4 Troubleshooting Doc 031842 01 5 02 4 9 Figure 4 1 ASE 100 Plumbing Diagram...
Page 106: ...ASE 100 Operator s Manual 5 36 Doc 031842 01 5 02...
Page 110: ...ASE 100 Operator s Manual A 4 Doc 031842 01 5 02...
Page 113: ...B Installation Doc 031842 01 5 02 B 3 Figure B 1 ASE 100 Rear Panel...
Page 150: ...ASE 100 Accelerated Solvent Extractor Index 8 Doc 031842 01 5 02...