ASE 100 Operator’s Manual
2-16
Doc. 031842-01 5/02
OPP (Organophosphorous Pesticides) Method Parameters
Solvent
MeCl
2
/Acetone (1:1, v/v)
Temperature
100
°
C
Static Time
5 min
Flush Volume
60%
Purge Time
100 sec
Static Cycle
1
PCB (Polychlorinated Biphenyls) Method Parameters
Solvent
Hexane
Temperature
100
°
C
Static Time
5 min
Flush Volume
60%
Purge Time
100 sec
Static Cycle
1
PDF (Dioxins and Furans) Method Parameters
Solvent
Toluene
Temperature
200
°
C
Static Time
5 min
Flush Volume
60%
Purge Time
100 sec
Static Cycle
1