C • User Interface
Doc. 031842-01 5/02
C-13
C.2.5
Extraction Counters Screen
Use the
EXTRACTION COUNTERS
screen to check the number of
extractions the system has run or the number of pump strokes performed.
This can be helpful in scheduling the routine replacement of certain parts.
Figure C-12. Extraction Counters Screen
Parameter
Description
EXTRACTION
Indicates the number of extractions performed since
the ASE 100 was installed or the extraction counter
was last reset.
PUMP STROKE
Indicates the number of pump strokes performed
since the ASE 100 was installed or the pump stroke
counter was last reset.
EXTR. RESET
Select
RESET
to reset the extraction counter to zero.
PUMP RESET
Select
RESET
to reset the pump stroke counter to
zero.