5 • Service
Doc. 031842-01 5/02
5-21
e.
Pull the trim panel completely off, exposing the inner front panel,
and set the trim panel aside.
6. Remove the right side panel of the ASE 100:
a.
Using a Phillips screwdriver, remove the two screws on the inner
front panel, near the needle assembly (see Figure 5-5).
b.
Insert a screwdriver into the hole located beneath the top screw
hole (see Figure 5-5). Push the screwdriver in until the right side
panel slides back about 1/2 inch. To prevent scratches or other
damage to the panel, do not let the panel fall onto the
workbench.
c.
Remove the right side panel from the system and set it aside.
7. Using a 1/4-inch open-end wrench, disconnect the stainless steel fitting on the
right side of the static valve (see Figure 5-8).
8. The needle assembly is secured to the inner front panel. The source needle
assembly is on the right side of the needle block (see Figure 5-9). Disconnect
the 1/4-inch stainless steel fitting (part of the source needle assembly) from
the needle block.
Figure 5-9. Vent and Source Needles
Summary of Contents for ASE 100
Page 8: ...ASE 100 Accelerated Solvent Extractor vi Doc 031842 01 5 02...
Page 24: ...ASE 100 Operator s Manual 2 10 Doc 031842 01 5 02 Figure 2 4 ASE 100 Rear Panel...
Page 32: ...ASE 100 Operator s Manual 2 18 Doc 031842 01 5 02...
Page 42: ...ASE 100 Operator s Manual 3 10 Doc 031842 01 5 02 Figure 3 2 Inspecting Cell O Rings...
Page 60: ...ASE 100 Operator s Manual 3 28 Doc 031842 01 5 02...
Page 69: ...4 Troubleshooting Doc 031842 01 5 02 4 9 Figure 4 1 ASE 100 Plumbing Diagram...
Page 106: ...ASE 100 Operator s Manual 5 36 Doc 031842 01 5 02...
Page 110: ...ASE 100 Operator s Manual A 4 Doc 031842 01 5 02...
Page 113: ...B Installation Doc 031842 01 5 02 B 3 Figure B 1 ASE 100 Rear Panel...
Page 150: ...ASE 100 Accelerated Solvent Extractor Index 8 Doc 031842 01 5 02...