B • Installation
Doc. 031842-01 5/02
B-15
B.2.8
Inspecting the Cell
O-rings are installed in the exterior ends of each extraction cell end cap,
as well as in the ends of the rinse cell. Before each run, inspect the O-
rings (see Figure B-10).
_
If an O-ring is dislodged: Press it into place, using the O-ring
insertion tool (P/N 049660).
_
If an O-ring is missing: Place a new O-ring over the opening in the
end of the cell cap or rinse cell and press the O-ring into place (see
Figure B-11), using the O-ring insertion tool (P/N 049660).
Figure B-10. Inspecting Cell O-Rings
Figure B-11. Installing Cell O-Rings