3 • Operation and Maintenance
Doc. 031842-01 5/02
3-3
3.1.2
Filling the Solvent Bottle
NOTE Never fill the solvent bottle or disconnect the tubing
connections to the
SOLVENT
and
GAS
connectors
(see Figure 3-1) when the system is performing an
extraction or a rinse cycle. During these times, the
solvent bottle is pressurized. If you remove the
bottle cap when the solvent bottle is pressurized, the
ASE 100 may not operate to specification.
1. Fill the solvent bottle with prepared solvent to the level indicated in
If you plan to run a preprogrammed method, refer to Section 2.5 for
the recommended solvent. If you plan to run a custom method, review
the solvent selection guidelines in Section 3.1.1.
NOTE The solvent level in the bottle must remain below the
gas inlet line. This prevents solvent from coming
into contact with the pneumatic valves.
2.
Place the bottle in the recess on top of the system.
Use only Dionex solvent bottles (1-liter, P/N 045900; 2-liter,
P/N 045901). These are glass bottles with a plastic,
shatterproof coating. Make sure the pressure applied to the
bottles does not exceed 0.07 MPa (10 psi).
Utilisez uniquement des réservoirs à solvant Dionex (1 litre,
N/P 045900; 2 litres, N/P 045901). Ce sont des réservoirs en
verre à revêtement incassable en plastique. Veillez à ce que la
pression exercée sur ces réservoirs ne dépasse pas 0,07 MPa.
Verwenden Sie ausschließlich die Lösemittelbehälter von
Dionex (1-Liter, Bestell-Nr. 045900; 2-Liter, Bestell-Nr. 045901).
Dabei handelt es sich um Glasbehälter mit einer
splittersicheren Plastikbeschichtung. Vergewissern Sie sich,
daß der Druck, der auf die Behälter ausgeübt wird, 0,07 MPa
nicht übersteigt.
Summary of Contents for ASE 100
Page 8: ...ASE 100 Accelerated Solvent Extractor vi Doc 031842 01 5 02...
Page 24: ...ASE 100 Operator s Manual 2 10 Doc 031842 01 5 02 Figure 2 4 ASE 100 Rear Panel...
Page 32: ...ASE 100 Operator s Manual 2 18 Doc 031842 01 5 02...
Page 42: ...ASE 100 Operator s Manual 3 10 Doc 031842 01 5 02 Figure 3 2 Inspecting Cell O Rings...
Page 60: ...ASE 100 Operator s Manual 3 28 Doc 031842 01 5 02...
Page 69: ...4 Troubleshooting Doc 031842 01 5 02 4 9 Figure 4 1 ASE 100 Plumbing Diagram...
Page 106: ...ASE 100 Operator s Manual 5 36 Doc 031842 01 5 02...
Page 110: ...ASE 100 Operator s Manual A 4 Doc 031842 01 5 02...
Page 113: ...B Installation Doc 031842 01 5 02 B 3 Figure B 1 ASE 100 Rear Panel...
Page 150: ...ASE 100 Accelerated Solvent Extractor Index 8 Doc 031842 01 5 02...