C • User Interface
Doc. 031842-01 5/02
C-11
C.2.3
Regulators Screen
Use the
REGULATORS
screen to monitor various pressure readings for the
ASE 100. You cannot edit any information on this screen.
NOTE Select the unit of measure for on-screen pressure
readings on the
SETUP
Figure C-10. Regulators Screen
Parameter
Description
SYSTEM
Indicates the pressure from the nitrogen gas source.
COMPRESSION
Indicates the nitrogen gas pressure applied to the
oven compression system. During an extraction, the
oven is compressed and the reading is approximately
1.03 MPa (130 psi). When the oven is not
compressed, this reading is zero.
BOTTLE
Indicates the pressure applied to the solvent bottle.
AUTOSEAL
Indicates the pressure applied to the AutoSeal air
cylinder. When the air cylinder is actuated, the
reading is 0.10
±
0.01 MPa (15
±
2 psi). When the air
cylinder is not actuated, this reading is zero.
Summary of Contents for ASE 100
Page 8: ...ASE 100 Accelerated Solvent Extractor vi Doc 031842 01 5 02...
Page 24: ...ASE 100 Operator s Manual 2 10 Doc 031842 01 5 02 Figure 2 4 ASE 100 Rear Panel...
Page 32: ...ASE 100 Operator s Manual 2 18 Doc 031842 01 5 02...
Page 42: ...ASE 100 Operator s Manual 3 10 Doc 031842 01 5 02 Figure 3 2 Inspecting Cell O Rings...
Page 60: ...ASE 100 Operator s Manual 3 28 Doc 031842 01 5 02...
Page 69: ...4 Troubleshooting Doc 031842 01 5 02 4 9 Figure 4 1 ASE 100 Plumbing Diagram...
Page 106: ...ASE 100 Operator s Manual 5 36 Doc 031842 01 5 02...
Page 110: ...ASE 100 Operator s Manual A 4 Doc 031842 01 5 02...
Page 113: ...B Installation Doc 031842 01 5 02 B 3 Figure B 1 ASE 100 Rear Panel...
Page 150: ...ASE 100 Accelerated Solvent Extractor Index 8 Doc 031842 01 5 02...