ASE 100 Operator’s Manual
3-12
Doc. 031842-01 5/02
3.1.7
Rinsing/Priming the System
During a rinse cycle, approximately 5 mL of solvent is pumped through
the system. Run a rinse cycle at the following times:
_
After initial setup
_
After refilling the solvent bottle
_
After changing solvents (rinse twice to remove all of the previous
solvent)
_
When the solvent lines contain bubbles
_
After a run is aborted
_
Before shutting down the system, if an extraction used acidic solvents
or basic solvents (see Section 3.8)
_
After a shutdown of more than one day (see Section 3.8)
Before starting a rinse cycle:
_
Check that the rinse cell size is matched to the extraction cell size.
_
Adjust the cell holder to accommodate the rinse cell, if necessary.
_
Install the rinse cell in the cell holder.
_
Close the cell door.
When you are ready, press the
RINSE
button to start the rinse cycle. For
detailed instructions, refer to Section B.2.6.
Summary of Contents for ASE 100
Page 8: ...ASE 100 Accelerated Solvent Extractor vi Doc 031842 01 5 02...
Page 24: ...ASE 100 Operator s Manual 2 10 Doc 031842 01 5 02 Figure 2 4 ASE 100 Rear Panel...
Page 32: ...ASE 100 Operator s Manual 2 18 Doc 031842 01 5 02...
Page 42: ...ASE 100 Operator s Manual 3 10 Doc 031842 01 5 02 Figure 3 2 Inspecting Cell O Rings...
Page 60: ...ASE 100 Operator s Manual 3 28 Doc 031842 01 5 02...
Page 69: ...4 Troubleshooting Doc 031842 01 5 02 4 9 Figure 4 1 ASE 100 Plumbing Diagram...
Page 106: ...ASE 100 Operator s Manual 5 36 Doc 031842 01 5 02...
Page 110: ...ASE 100 Operator s Manual A 4 Doc 031842 01 5 02...
Page 113: ...B Installation Doc 031842 01 5 02 B 3 Figure B 1 ASE 100 Rear Panel...
Page 150: ...ASE 100 Accelerated Solvent Extractor Index 8 Doc 031842 01 5 02...