ASE 100 Operator’s Manual
A-2
Doc. 031842-01 5/02
A.4
Pneumatic
A.5
Front Panel Display and Keypad
A.6
Extraction Cells
A.7
Collection Bottles
A.8
Interior Components
Nitrogen
1.03 to 1.38 MPa (150 to 200 psi)
Display
Vacuum fluorescence display (VFD)
Keypad
Seven buttons for entering commands and selecting screen
parameters
Extraction Cells
Stainless steel cell bodies and end caps with seals and stainless
steel frits. Cells are available in four sizes: 10 mL (10.0 mL
actual volume), 34 mL (33.0 mL actual volume), 66 mL (66.0
mL actual volume), and 100 mL (99.8 mL actual volume). The
internal diameter of the 10 mL cell is 16 mm (0.61 in). The
internal diameter of the other cells is 28 mm (1.12 in).
Collection Bottles
250 mL bottle; the bottle septum is Teflon-coated on the solvent
side
Oven
Heats from 40 to 200 °C; the oven turns off automatically after
eight hours of no system activity
Pump
Operating pressure of 10.35 MPa (1500 psi)
Valves
High pressure valves: purge, pressure relief, and static
Bottle Sensors
When the needles are in the “down” position, sensors detect
whether a collection bottle is present and whether it is full
Summary of Contents for ASE 100
Page 8: ...ASE 100 Accelerated Solvent Extractor vi Doc 031842 01 5 02...
Page 24: ...ASE 100 Operator s Manual 2 10 Doc 031842 01 5 02 Figure 2 4 ASE 100 Rear Panel...
Page 32: ...ASE 100 Operator s Manual 2 18 Doc 031842 01 5 02...
Page 42: ...ASE 100 Operator s Manual 3 10 Doc 031842 01 5 02 Figure 3 2 Inspecting Cell O Rings...
Page 60: ...ASE 100 Operator s Manual 3 28 Doc 031842 01 5 02...
Page 69: ...4 Troubleshooting Doc 031842 01 5 02 4 9 Figure 4 1 ASE 100 Plumbing Diagram...
Page 106: ...ASE 100 Operator s Manual 5 36 Doc 031842 01 5 02...
Page 110: ...ASE 100 Operator s Manual A 4 Doc 031842 01 5 02...
Page 113: ...B Installation Doc 031842 01 5 02 B 3 Figure B 1 ASE 100 Rear Panel...
Page 150: ...ASE 100 Accelerated Solvent Extractor Index 8 Doc 031842 01 5 02...