5 • Service
Doc. 031842-01 5/02
5-7
6. Remove the right side panel of the ASE 100:
a.
Using a Phillips screwdriver, remove the two screws on the inner
front panel, near the needle assembly (see Figure 5-5).
b.
Insert a screwdriver into the hole located beneath the top screw
hole (see Figure 5-5). Push the screwdriver in until the right side
panel slides back about 1/2 inch. To prevent scratches or other
damage to the panel, do not let the panel fall onto the
workbench.
c.
Remove the right side panel from the system and set it aside.
To reinstall the right side panel:
1. Hold the panel against the right side of the system. Engage the small tabs on
the top and bottom of the panel with the slots in the chassis. Pull the panel
forward until it locks into place.
2. Replace the two Phillips screws on the inner front panel.
3. Carefully align the lower trim panel with the edges of the instrument, and then
gently push the panel into place.
4. Reinstall the waste bottle.
5. Turn on the ASE 100 main power switch.
6. Reconnect the gas source at the ASE 100 rear panel and turn on the gas
supply.
Figure 5-5. Removing the Right Side Panel
Summary of Contents for ASE 100
Page 8: ...ASE 100 Accelerated Solvent Extractor vi Doc 031842 01 5 02...
Page 24: ...ASE 100 Operator s Manual 2 10 Doc 031842 01 5 02 Figure 2 4 ASE 100 Rear Panel...
Page 32: ...ASE 100 Operator s Manual 2 18 Doc 031842 01 5 02...
Page 42: ...ASE 100 Operator s Manual 3 10 Doc 031842 01 5 02 Figure 3 2 Inspecting Cell O Rings...
Page 60: ...ASE 100 Operator s Manual 3 28 Doc 031842 01 5 02...
Page 69: ...4 Troubleshooting Doc 031842 01 5 02 4 9 Figure 4 1 ASE 100 Plumbing Diagram...
Page 106: ...ASE 100 Operator s Manual 5 36 Doc 031842 01 5 02...
Page 110: ...ASE 100 Operator s Manual A 4 Doc 031842 01 5 02...
Page 113: ...B Installation Doc 031842 01 5 02 B 3 Figure B 1 ASE 100 Rear Panel...
Page 150: ...ASE 100 Accelerated Solvent Extractor Index 8 Doc 031842 01 5 02...