5 • Service
Doc. 031842-01 5/02
5-19
7. The static valve is installed in the lower left corner of the component panel
(see Figure 5-8). Disconnect the orange tubing from the elbow press fitting on
the valve.
8. Disconnect the stainless steel line from the right side of the valve. (This line
connects to the source needle.)
9. Disconnect the stainless steel line (P/N 059821) from the top of the valve.
(This line connects to the lower AutoSeal tip.)
10. Loosen the Allen hex screw (9/64-inch) on the clamp ring below the valve
base.
11. Lift the old valve off the valve base.
12. Lower the new static valve (P/N 048778) onto the valve base. Tighten the
Allen hex screw on the clamp ring.
13. Reconnect all of the lines.
Figure 5-8. Connections to the Static Valve