C • User Interface
Doc. 031842-01 5/02
C-3
C.1.2
Status Screen
Use the
STATUS
screen to monitor the progress of a method run. The
screen is updated three times per second. You cannot edit any information
on this screen.
Figure C-3. Status Screen
Parameter
Description
STATUS
The method step in progress, if any, or the current
system status. If the latter is indicated, the
STATUS
field
will display one of the following descriptions: Initialize,
Idle, Load, Fill, Preheat, Heat, Oven Wait, Oven Ready,
Static, Flush, Purge, Relief, Unload, Rinse, or Abort. If
the status is Static or Flush, the cycle number is
indicated.
METHOD
The name or number of the current method.
VOL
The approximate volume (in mL) of solvent delivered
by the pump since the method started.
PRESSURE
The current pressure reading. Select the unit of measure
on the
SETUP
screen (see Section C.1.3).
TEMP
The oven temperature specified in the method.
TIME
Elapsed time (in minutes and seconds) since the method
started running.
Summary of Contents for ASE 100
Page 8: ...ASE 100 Accelerated Solvent Extractor vi Doc 031842 01 5 02...
Page 24: ...ASE 100 Operator s Manual 2 10 Doc 031842 01 5 02 Figure 2 4 ASE 100 Rear Panel...
Page 32: ...ASE 100 Operator s Manual 2 18 Doc 031842 01 5 02...
Page 42: ...ASE 100 Operator s Manual 3 10 Doc 031842 01 5 02 Figure 3 2 Inspecting Cell O Rings...
Page 60: ...ASE 100 Operator s Manual 3 28 Doc 031842 01 5 02...
Page 69: ...4 Troubleshooting Doc 031842 01 5 02 4 9 Figure 4 1 ASE 100 Plumbing Diagram...
Page 106: ...ASE 100 Operator s Manual 5 36 Doc 031842 01 5 02...
Page 110: ...ASE 100 Operator s Manual A 4 Doc 031842 01 5 02...
Page 113: ...B Installation Doc 031842 01 5 02 B 3 Figure B 1 ASE 100 Rear Panel...
Page 150: ...ASE 100 Accelerated Solvent Extractor Index 8 Doc 031842 01 5 02...