3 • Operation and Maintenance
Doc. 031842-01 5/02
3-27
scratched, gouged, or corroded. If the tip is damaged or worn, replace
it (see Section 5.10).
3.8
Shutdown
Overnight Shutdown
_
After extracting with a 100% organic solvent, you may turn off the power
immediately. No other action is required.
_
After extracting with an acidic solvents or a basic solvent, rinse the system
with either 100% organic solvent or distilled water (see Section 3.1.7) before
turning off the power.
Longer Shutdowns
_
Before shutting down for more than one night:
1. After extracting with acidic solvents or basic solvents, rinse the
system with either 100% organic solvent or distilled water (see
Section 3.1.7) before turning off the power.
2. Turn off the gas supply.
_
Before shipping the ASE 100:
1. After extracting with acidic solvents or basic solvents, rinse the
system with either 100% organic solvent (e.g., acetone or methanol)
or distilled water (see Section 3.1.7).
2. Empty the solvent bottle, reconnect the bottle to the system, and run
one or more rinse cycles to ensure that all solvent is removed from the
lines.
3. Remove the solvent bottle.
Summary of Contents for ASE 100
Page 8: ...ASE 100 Accelerated Solvent Extractor vi Doc 031842 01 5 02...
Page 24: ...ASE 100 Operator s Manual 2 10 Doc 031842 01 5 02 Figure 2 4 ASE 100 Rear Panel...
Page 32: ...ASE 100 Operator s Manual 2 18 Doc 031842 01 5 02...
Page 42: ...ASE 100 Operator s Manual 3 10 Doc 031842 01 5 02 Figure 3 2 Inspecting Cell O Rings...
Page 60: ...ASE 100 Operator s Manual 3 28 Doc 031842 01 5 02...
Page 69: ...4 Troubleshooting Doc 031842 01 5 02 4 9 Figure 4 1 ASE 100 Plumbing Diagram...
Page 106: ...ASE 100 Operator s Manual 5 36 Doc 031842 01 5 02...
Page 110: ...ASE 100 Operator s Manual A 4 Doc 031842 01 5 02...
Page 113: ...B Installation Doc 031842 01 5 02 B 3 Figure B 1 ASE 100 Rear Panel...
Page 150: ...ASE 100 Accelerated Solvent Extractor Index 8 Doc 031842 01 5 02...