ASE 100 Operator’s Manual
3-6
Doc. 031842-01 5/02
Sample Preparation Guidelines
The mixtures here are recommendations; adjust the proportions as
required for your extraction.
_
If the sample appears dry, use one of these mixtures:
4 grams sample to 1 gram DE
4 grams sample to 4 grams Na
2
SO
4
Mix the sample and the DE or Na
2
SO
4
thoroughly in a small bottle,
beaker, or mortar.
_
If the sample appears wet, use one of these mixtures:
4 grams sample to 2 grams DE
4 grams sample to 8 grams Na
2
SO
4
Mix the sample and the DE or Na
2
SO
4
thoroughly in a small bottle,
beaker, or mortar.
_
If the sample is pure liquid, use 5 grams sample to 3 grams DE.
Fill the cell with DE and then add the sample.
Summary of Contents for ASE 100
Page 8: ...ASE 100 Accelerated Solvent Extractor vi Doc 031842 01 5 02...
Page 24: ...ASE 100 Operator s Manual 2 10 Doc 031842 01 5 02 Figure 2 4 ASE 100 Rear Panel...
Page 32: ...ASE 100 Operator s Manual 2 18 Doc 031842 01 5 02...
Page 42: ...ASE 100 Operator s Manual 3 10 Doc 031842 01 5 02 Figure 3 2 Inspecting Cell O Rings...
Page 60: ...ASE 100 Operator s Manual 3 28 Doc 031842 01 5 02...
Page 69: ...4 Troubleshooting Doc 031842 01 5 02 4 9 Figure 4 1 ASE 100 Plumbing Diagram...
Page 106: ...ASE 100 Operator s Manual 5 36 Doc 031842 01 5 02...
Page 110: ...ASE 100 Operator s Manual A 4 Doc 031842 01 5 02...
Page 113: ...B Installation Doc 031842 01 5 02 B 3 Figure B 1 ASE 100 Rear Panel...
Page 150: ...ASE 100 Accelerated Solvent Extractor Index 8 Doc 031842 01 5 02...