ASE 100 Operator’s Manual
5-4
Doc. 031842-01 5/02
5.2
Replacing the Cell O-Ring
1. Use the small flathead screwdriver (P/N 046985) provided in the ASE 100
Ship Kit (P/N 059397) to remove worn O-rings. To do so, insert the tip of the
screwdriver into the end cap and carefully pick out the O-ring.
2. Place a new O-ring (Teflon: P/N 049457, pkg. of 50; Viton: P/N 056325, pkg.
of 50) over the opening in the end of the cell cap and press it into place, using
the O-ring insertion tool (P/N/ 049660) (see Figure 5-2).
Be careful not to scratch the interior of the cell end cap when
installing or removing the cell O-ring. Scratches on the sealing
surface will prevent the O-ring from sealing properly and may
result in leaks during operation.
Figure 5-2. Installing the Cell O-Ring