CyberOptics Corporation
37
Connecting the Air Supply
Connecting the Air Supply
The SE500
ULTRA
requires a connection to clean, dry compressed air at a pressure of 90 ± 10 psi
(620 ± 69KPa).
➢
To connect air supply to SE500
ULTRA
1. Make sure that the air line to the SE500
ULTRA
is unpressurized.
2. Push the air hose connector onto the air regulator quick-connect fitting. Air tubing size
should be 6 mm (outside diameter) x 4 mm (inside diameter).
3. Turn on power to the system.
4. Turn the knob at the top of the air regulator to set the air pressure to 90 ± 10 psi
(620 ± 69KPa).
5. Observe the air pressure on the gauge located outside the machine near the E-Stop button.
Figure 23.
Connecting to the Air Regulator
Turn this knob to
set air pressure
Quick-connect fitting
Back Electrical Panel
i
Note:
The default setting for the air pressure gauge should display the air pressure in psi. If not,
or to change other settings on the gauge, refer to “