Operator’s Manual 4-35
110017276EN Rev. B
4 Detailed Reference
Description of Ultrasound Modes
The application of ultrasound power may be fixed or linear. Linear power is proportionally controlled by the
footpedal between zero and the maximum limit set on the console.
Ultrasound power may be adjusted from 0% to 100% in 5% increments using the up/down spin buttons, Foot
Control buttons, or remote, and 1% increments using the keypad. The ultrasound output will be activated at the
minimum programmed power level as the footpedal moves into the active ultrasound region, and will increase to
the maximum programmed output as a function of linear footpedal travel.
Single Linear Ultrasound Mode with Fixed Aspiration
Once irrigation has been initiated and the footpedal has been depressed approximately five degrees (or as
programmed), an increase in footpedal resistance will be noted signifying the transition from Region 1 to
Region 2 and the start of aspiration. Fixed aspiration will be developed at the selected aspiration level. The
screen will display the actual amount of aspiration at any given time.
Ultrasound power is activated in Region 3 of footpedal travel. Another momentary increase in footpedal
resistance will be noted (if detents are enabled) signifying the transition from one Region to the next, and the
start of ultrasound power. Ultrasound power will be initiated and controlled as a function of footpedal travel in
Region 3. The next ultrasound submode may be selected (if enabled) by moving the footpedal in the outward
yaw direction.
Single Linear Ultrasound Mode with Linear Aspiration
Once irrigation has been initiated and the footpedal has been depressed approximately five degrees (or as
programmed), an increase in footpedal resistance will be noted signifying the transition from Region 1 to
Region 2 and the start of aspiration. Aspiration will increase from 0 to the fixed level in proportion to footpedal
travel in Region 2. The screen will display the actual amount of aspiration. Aspiration will remain at the fixed
level in Region 3.
Ultrasound power is activated in Region 3 of footpedal travel. Another momentary increase in footpedal
resistance will be noted signifying the transition from one region to the next, and the start of ultrasound power.
Linear ultrasound power will be initiated and controlled as a linear function of footpedal travel in Region 3.
Pulsed ultrasound may be toggled on/off by moving the footpedal in the outward yaw direction.
Note:
If single or multiple burst mode is selected, position 3 (or outward yaw movement) does not
control ultrasound power, but rather the burst interval (for multiple burst) or nearly full travel
initiates and resets the single burst (see page 4-36 ).
Summary of Contents for STELLARIS PC
Page 1: ...Operator s Manual ...
Page 78: ...1 68 Operator s Manual 110017276EN Rev B 1 Getting Started ...
Page 190: ...3 36 Operator s Manual 110017276EN Rev B 3 ...
Page 244: ...4 54 Operator s Manual 110017276EN Rev B 4 ...
Page 284: ...6 26 Operator s Manual 110017276EN Rev B 6 Troubleshooting ...
Page 329: ...Preface ...
Page 330: ...Table of Contents ...
Page 331: ...Getting Started ...
Page 332: ...User Interface ...
Page 333: ...Customizing Your System ...
Page 334: ...Detailed Reference ...
Page 335: ...Cleaning and Sterilization ...
Page 336: ...Troubleshooting ...
Page 337: ...Service and Warranty ...
Page 338: ...Specifications ...
Page 339: ...Glossary ...
Page 340: ...Index ...
Page 341: ...Preface ...
Page 342: ...Table of Contents ...
Page 343: ...Getting Started ...
Page 344: ...User Interface ...
Page 345: ...Customizing Your System ...
Page 346: ...Detailed Reference ...
Page 347: ...Cleaning and Sterilization ...
Page 348: ...Troubleshooting ...
Page 349: ...Service and Warranty ...
Page 350: ...Specifications ...
Page 351: ...Glossary ...
Page 352: ...Index ...