1-22 Operator’s Manual
110017276EN Rev. B
1 Getting Started
Figure 1.12.
Prime and Tune Screen
. This is an example of a posterior domain screen.
5. Advance to Surgery Phase
WARNING:
Inadvertent activation of functions that are intended for priming or tuning handpieces while
the handpiece is in the eye can create a hazardous situation that could result in patient injury.
Once the system has been successfully primed and tuned, it will automatically move to the main surgical screen.
Manually selecting
Advance to Surgery
produces the same result.
Note:
If the system is not primed and tuned, the aspiration, vitrectomy and ultrasound
functions will be unavailable.
Summary of Contents for STELLARIS PC
Page 1: ...Operator s Manual ...
Page 78: ...1 68 Operator s Manual 110017276EN Rev B 1 Getting Started ...
Page 190: ...3 36 Operator s Manual 110017276EN Rev B 3 ...
Page 244: ...4 54 Operator s Manual 110017276EN Rev B 4 ...
Page 284: ...6 26 Operator s Manual 110017276EN Rev B 6 Troubleshooting ...
Page 329: ...Preface ...
Page 330: ...Table of Contents ...
Page 331: ...Getting Started ...
Page 332: ...User Interface ...
Page 333: ...Customizing Your System ...
Page 334: ...Detailed Reference ...
Page 335: ...Cleaning and Sterilization ...
Page 336: ...Troubleshooting ...
Page 337: ...Service and Warranty ...
Page 338: ...Specifications ...
Page 339: ...Glossary ...
Page 340: ...Index ...
Page 341: ...Preface ...
Page 342: ...Table of Contents ...
Page 343: ...Getting Started ...
Page 344: ...User Interface ...
Page 345: ...Customizing Your System ...
Page 346: ...Detailed Reference ...
Page 347: ...Cleaning and Sterilization ...
Page 348: ...Troubleshooting ...
Page 349: ...Service and Warranty ...
Page 350: ...Specifications ...
Page 351: ...Glossary ...
Page 352: ...Index ...