6-12 Operator’s Manual
110017276EN Rev. B
6 Troubleshooting
ID Type
Text Message
Suggested Action(s)
USM08
Info
The ultrasound handpiece has failed the
tuning process.
SG52 Ensure ultrasound needle is properly
tightened. Select ‘Prime / Tune’ or ‘Tune
Only’ again to repeat the handpiece tuning
process.
SG53 Unplug the handpiece and try another
ultrasound handpiece.
SG04 Call your product service
representative if this problem persists.
USM09
Info
The ultrasound handpiece may have
failed or may be failing.
SG54 Confirm proper ultrasound handpiece
operation in a test chamber.
SG53 Unplug the handpiece and try another
ultrasound handpiece.
SG04 Call your product service
representative if this problem persists.
USM10
Warning
Coagulation circuit may have failed,
potential coagulation output over-voltage
condition.
SG55 Confirm proper operation of the
coagulation handpiece in a beaker of BSS;
activate the function and ensure the error
message does appear again. If similar
message appears, stop using the function
and call B&L service support.
SG56 Unplug the handpiece and try another
bipolar coagulation cord.
SG04 Call your product service
representative if this problem persists.
USM11
Warning
Coagulation circuit may have failed,
potential uncommanded or incorrect
coagulation output.
SG55 Confirm proper operation of the
coagulation handpiece in a beaker of BSS;
activate the function and ensure the error
message does appear again. If similar
message appears, stop using the function
and call B&L service support.
SG56 Unplug the handpiece and try another
bipolar coagulation cord.
SG04 Call your product service
representative if this problem persists.
USM15
Info
Unable to read ultrasound handpiece
data. Attempt to tune the handpiece to
confirm proper operation.
SG53 Unplug the handpiece and try another
ultrasound handpiece.
SG04 Call your product service
representative if this problem persists.
Summary of Contents for STELLARIS PC
Page 1: ...Operator s Manual ...
Page 78: ...1 68 Operator s Manual 110017276EN Rev B 1 Getting Started ...
Page 190: ...3 36 Operator s Manual 110017276EN Rev B 3 ...
Page 244: ...4 54 Operator s Manual 110017276EN Rev B 4 ...
Page 284: ...6 26 Operator s Manual 110017276EN Rev B 6 Troubleshooting ...
Page 329: ...Preface ...
Page 330: ...Table of Contents ...
Page 331: ...Getting Started ...
Page 332: ...User Interface ...
Page 333: ...Customizing Your System ...
Page 334: ...Detailed Reference ...
Page 335: ...Cleaning and Sterilization ...
Page 336: ...Troubleshooting ...
Page 337: ...Service and Warranty ...
Page 338: ...Specifications ...
Page 339: ...Glossary ...
Page 340: ...Index ...
Page 341: ...Preface ...
Page 342: ...Table of Contents ...
Page 343: ...Getting Started ...
Page 344: ...User Interface ...
Page 345: ...Customizing Your System ...
Page 346: ...Detailed Reference ...
Page 347: ...Cleaning and Sterilization ...
Page 348: ...Troubleshooting ...
Page 349: ...Service and Warranty ...
Page 350: ...Specifications ...
Page 351: ...Glossary ...
Page 352: ...Index ...