5-2 Operator’s Manual
110017276EN Rev. B
5 Cleaning and Sterilization Requirements
You should clean the fluid level detection lens ( Figure 5.1 ) every three months with a 4 x 4 in. gauze pad and
isopropyl alcohol.
Figure 5.1. Fluid Level Detection Lens.
1. Window.
Gently swab the electrical connectors with an alcohol swab weekly, taking care to avoid excessive quantities of
cleaning solution around the ports. Do not reconnect to power until the ports have completely dried.
5.2.
Bipolar Coagulation Accessories
WARNING:
The coagulation accessories should not be sterilized using a cold soaking solution.
WARNING:
All bipolar accessories must be rated for an operating voltage of at least 120V.
CAUTION:
Allow 20 minutes after sterilization for the handpiece and cord to cool before using them
again. The handpiece connector must be completely dry before it is connected it to equipment.
Summary of Contents for STELLARIS PC
Page 1: ...Operator s Manual ...
Page 78: ...1 68 Operator s Manual 110017276EN Rev B 1 Getting Started ...
Page 190: ...3 36 Operator s Manual 110017276EN Rev B 3 ...
Page 244: ...4 54 Operator s Manual 110017276EN Rev B 4 ...
Page 284: ...6 26 Operator s Manual 110017276EN Rev B 6 Troubleshooting ...
Page 329: ...Preface ...
Page 330: ...Table of Contents ...
Page 331: ...Getting Started ...
Page 332: ...User Interface ...
Page 333: ...Customizing Your System ...
Page 334: ...Detailed Reference ...
Page 335: ...Cleaning and Sterilization ...
Page 336: ...Troubleshooting ...
Page 337: ...Service and Warranty ...
Page 338: ...Specifications ...
Page 339: ...Glossary ...
Page 340: ...Index ...
Page 341: ...Preface ...
Page 342: ...Table of Contents ...
Page 343: ...Getting Started ...
Page 344: ...User Interface ...
Page 345: ...Customizing Your System ...
Page 346: ...Detailed Reference ...
Page 347: ...Cleaning and Sterilization ...
Page 348: ...Troubleshooting ...
Page 349: ...Service and Warranty ...
Page 350: ...Specifications ...
Page 351: ...Glossary ...
Page 352: ...Index ...