6-4 Operator’s Manual
110017276EN Rev. B
6 Troubleshooting
Power Supply Module Messages
ID Type
Text Message
Suggested Action(s)
BPS02
Warning
The power supply was not detected in the
system.
SG01 Call your product service
representative.
BPS03
Warning
The power supply software version is not
compatible with this software version.
Surgical mode is not available.
SG02 Call your product service
representative. A compatible software
version must be downloaded.
BPS04
Warning
The power supply has failed to respond
to a settings command. The module
settings have been re-sent to the module.
SG04 Call your product service
representative if this problem persists.
BPS05
Warning
The power supply has reset. The module
settings have been re-sent to the module.
SG04 Call your product service
representative if this problem persists.
BPS08
Warning
The power module is at an over
temperature condition. System shutdown
is imminent.
SG67 Confirm that the bottom and lower
rear areas of the system are not blocked
from free air flow.
SG04 Call your product service
representative if this problem persists.
Compressor Module Messages
ID Type
Text Message
Suggested Action(s)
CPX01
Warning
The compressor module was not detected
in the system. Surgical mode is not
available.
SG01 Call your product service
representative.
CPX02
Warning
The compressor module software version
is not compatible with this software
version. Surgical mode is not available.
SG02 Call your product service
representative. A compatible software
version must be downloaded.
CPX03
Warning
The compressor module has failed to
respond to a settings command. The
module settings have been re-sent to the
module.
SG04 Call your product service
representative if this problem persists.
CPX04
Warning
The compressor module has reset. The
module settings have been re-sent to the
module.
SG04 Call your product service
representative if this problem persists.
CPX05
Warning
The air pressure output is lower than
commanded.
SG44 Check that the air line tubing is
connected tightly to the system.
SG45 Check that there are no leaks in
the air line and that the air line tubing
is connected tightly to the other pack
components.
SG04 Call your product service
representative if this problem persists.
Summary of Contents for STELLARIS PC
Page 1: ...Operator s Manual ...
Page 78: ...1 68 Operator s Manual 110017276EN Rev B 1 Getting Started ...
Page 190: ...3 36 Operator s Manual 110017276EN Rev B 3 ...
Page 244: ...4 54 Operator s Manual 110017276EN Rev B 4 ...
Page 284: ...6 26 Operator s Manual 110017276EN Rev B 6 Troubleshooting ...
Page 329: ...Preface ...
Page 330: ...Table of Contents ...
Page 331: ...Getting Started ...
Page 332: ...User Interface ...
Page 333: ...Customizing Your System ...
Page 334: ...Detailed Reference ...
Page 335: ...Cleaning and Sterilization ...
Page 336: ...Troubleshooting ...
Page 337: ...Service and Warranty ...
Page 338: ...Specifications ...
Page 339: ...Glossary ...
Page 340: ...Index ...
Page 341: ...Preface ...
Page 342: ...Table of Contents ...
Page 343: ...Getting Started ...
Page 344: ...User Interface ...
Page 345: ...Customizing Your System ...
Page 346: ...Detailed Reference ...
Page 347: ...Cleaning and Sterilization ...
Page 348: ...Troubleshooting ...
Page 349: ...Service and Warranty ...
Page 350: ...Specifications ...
Page 351: ...Glossary ...
Page 352: ...Index ...