![Bausch & Lomb STELLARIS PC Operator'S Manual Download Page 264](http://html.mh-extra.com/html/bausch-and-lomb/stellaris-pc/stellaris-pc_operators-manual_2711462264.webp)
6-6 Operator’s Manual
110017276EN Rev. B
6 Troubleshooting
ID Type
Text Message
Suggested Action(s)
CPX22
Warning
The internal air pressure is higher than
expected.
Ensure system input pressure is 72.5 to
100 psi. (5.0 bar - 6.9 bar).
SG25 Confirm correct infusion and
aspiration function.
SG31 Confirm correct vitrectomy function.
SG04 Call your product service
representative if this problem persists.
CPX23 Info
Source pressure is high.
SG102 Decrease input pressure to 85 psi.
IV Pole Module Messages
ID Type
Text Message
Suggested Action(s)
EIV01
Warning
The IV pole controller was not detected
in the system. The motorized IV pole
function is not available.
SG01 Call your product service
representative.
EIV02
Warning
The IV pole controller software version
is not compatible with this software
version. The motorized IV Pole function
is not available.
SG02 Call your product service
representative. A compatible software
version must be downloaded.
EIV03
Warning
The IV Pole controller has failed to
respond to a settings command. The
module settings have been re-sent to the
module.
SG04 Call your product service
representative if this problem persists.
EIV04
Warning
The IV pole controller has reset. The
module settings have been re-sent to the
module.
SG04 Call your product service
representative if this problem persists.
EIV05
Warning
The IV pole position cannot be
determined.
SG62 Please command the IV pole to the
minimum bottle height position.
SG04 Call your product service
representative if this problem persists.
EIV07
Warning
The IV pole is not detecting the home
position switch or the IV pole motor may
have failed.
SG62 Please command the IV pole to the
minimum bottle height position.
SG04 Call your product service
representative if this problem persists.
EIV08
Warning
The IV pole panel button sensors have
failed or the buttons have been activated
continuously since being powered on.
SG64 Ensure that the IV pole back panel
buttons are not activated.
SG65 Confirm proper operation of the IV
pole back panel buttons.
SG04 Call your product service
representative if this problem persists.
Summary of Contents for STELLARIS PC
Page 1: ...Operator s Manual ...
Page 78: ...1 68 Operator s Manual 110017276EN Rev B 1 Getting Started ...
Page 190: ...3 36 Operator s Manual 110017276EN Rev B 3 ...
Page 244: ...4 54 Operator s Manual 110017276EN Rev B 4 ...
Page 284: ...6 26 Operator s Manual 110017276EN Rev B 6 Troubleshooting ...
Page 329: ...Preface ...
Page 330: ...Table of Contents ...
Page 331: ...Getting Started ...
Page 332: ...User Interface ...
Page 333: ...Customizing Your System ...
Page 334: ...Detailed Reference ...
Page 335: ...Cleaning and Sterilization ...
Page 336: ...Troubleshooting ...
Page 337: ...Service and Warranty ...
Page 338: ...Specifications ...
Page 339: ...Glossary ...
Page 340: ...Index ...
Page 341: ...Preface ...
Page 342: ...Table of Contents ...
Page 343: ...Getting Started ...
Page 344: ...User Interface ...
Page 345: ...Customizing Your System ...
Page 346: ...Detailed Reference ...
Page 347: ...Cleaning and Sterilization ...
Page 348: ...Troubleshooting ...
Page 349: ...Service and Warranty ...
Page 350: ...Specifications ...
Page 351: ...Glossary ...
Page 352: ...Index ...