Operator’s Manual 6-15
110017276EN Rev. B
6 Troubleshooting
Foot Control Module Message
ID Type
Text Message
Suggested Action(s)
WFC01
Info
The Foot Control battery has exceeded
300 charge cycles. The Foot Control
battery may not provide power for the
entire day.
SG32 If the battery does not provide power
for the full day, replace the battery.
SG33 Use the wired cable connection to the
system.
SG34 Replace the Foot Control with
another Foot Control using the wired cable
connection to the system.
WFC02
Warning
The Foot Control does not have valid
factory calibration data. Surgical mode is
not available.
SG34 Replace the Foot Control with
another Foot Control using the wired cable
connection to the system.
SG35 Call your product service
representative to have the Foot Control
calibrated.
WFC03
Info
The Foot Control is programmed for
right footed operation, but is set up for
left footed operation.
SG36 Change the Foot Control home
position switch selection.
SG37 Confirm that the correct surgeon
settings are loaded. If incorrect, select
‘Setup’, then select ‘Select Surgeon’ to
select the correct surgeon settings.
WFC04
Info
The Foot Control is programmed for left
footed operation, but is set up for right
footed operation.
SG36 Change the Foot Control home
position switch selection.
SG37 Confirm that the correct surgeon
settings are loaded. If incorrect, select
‘Setup’, then select ‘Select Surgeon’ to
select the correct surgeon settings.
WFC05
Warning
The Foot Control center pedal or switch
sensors have failed or the Foot Control
has been activated continuously since
being powered on.
SG38 Reposition the Foot Control and
ensure that the center pedal and buttons are
not activated.
SG34 Replace the Foot Control with
another Foot Control using the wired cable
connection to the system.
SG04 Call your product service
representative if this problem persists.
WFC08
Info
System not detecting the Foot Control.
SG40 Initiate wireless Foot Control
connectivity by pressing one of the Foot
Control buttons momentarily, the left LED
will light up.
SG33 Use the wired cable connection to the
system.
SG34 Replace the Foot Control with
another Foot Control using the wired cable
connection to the system.
Summary of Contents for STELLARIS PC
Page 1: ...Operator s Manual ...
Page 78: ...1 68 Operator s Manual 110017276EN Rev B 1 Getting Started ...
Page 190: ...3 36 Operator s Manual 110017276EN Rev B 3 ...
Page 244: ...4 54 Operator s Manual 110017276EN Rev B 4 ...
Page 284: ...6 26 Operator s Manual 110017276EN Rev B 6 Troubleshooting ...
Page 329: ...Preface ...
Page 330: ...Table of Contents ...
Page 331: ...Getting Started ...
Page 332: ...User Interface ...
Page 333: ...Customizing Your System ...
Page 334: ...Detailed Reference ...
Page 335: ...Cleaning and Sterilization ...
Page 336: ...Troubleshooting ...
Page 337: ...Service and Warranty ...
Page 338: ...Specifications ...
Page 339: ...Glossary ...
Page 340: ...Index ...
Page 341: ...Preface ...
Page 342: ...Table of Contents ...
Page 343: ...Getting Started ...
Page 344: ...User Interface ...
Page 345: ...Customizing Your System ...
Page 346: ...Detailed Reference ...
Page 347: ...Cleaning and Sterilization ...
Page 348: ...Troubleshooting ...
Page 349: ...Service and Warranty ...
Page 350: ...Specifications ...
Page 351: ...Glossary ...
Page 352: ...Index ...