4-42 Operator’s Manual
110017276EN Rev. B
4 Detailed Reference
Foot Control of Anterior Vitrectomy Mode
As the footpedal is initially pressed, the irrigation control valve will open to allow irrigation into the eye.
Once irrigation has been initiated and the footpedal has been depressed approximately five degrees (or as
programmed), a momentary increase in footpedal resistance will be noted signifying the transition from
Region 1 to Region 2 and the start of aspiration. Aspiration increases proportional to footpedal travel with the
maximum level being set via the
Max Vacuum
input on the touch screen. Region 2 will provide linear control
of aspiration. The
Actual Vacuum
display will indicate the current aspiration level.
If enabled, an audible linear tone will indicate aspiration. The pitch of the tone increases with increased
aspiration.
For Fixed Cut vitrectomy mode, vitreous cutting is activated in Region 2. The cut rate is fixed. Each successive
outward yaw movement toggles the cutter ON or OFF. If enabled, an audible tone will indicate cutter on (double
beep) or cutter off (single beep).
For Dual Linear Cut mode, outward yaw movement provides linear control of the cut rate as a function of
footpedal displacement once it is within Region 2. The actual cut rate is displayed on the screen. When the
footpedal is released, it returns to center and the cutter is disabled. If enabled, an audible linear tone indicates
cut rate, and the pitch of the tone increases with increased cutter speed.
Reflux (if enabled) is by inward yaw movement of the footpedal.
By default, the vitrectomy cutter is On.
Pressurized Infusion Function
WARNING:
This system is to be used in either air pressured infusion mode or IV Pole mode but never
both together as high intraocular pressures may result.
Note:
Specific instruction for cleaning and sterilization included with the handpiece or accessory take
precedence over these instructions.
Note:
Once pressurized infusion pump is turned ON, it will continue to operate even when the function
is removed from the screen display.
Note:
When using air pressured infusion, hang the bottle so that the drip chamber is close to patient
eye level.
Summary of Contents for STELLARIS PC
Page 1: ...Operator s Manual ...
Page 78: ...1 68 Operator s Manual 110017276EN Rev B 1 Getting Started ...
Page 190: ...3 36 Operator s Manual 110017276EN Rev B 3 ...
Page 244: ...4 54 Operator s Manual 110017276EN Rev B 4 ...
Page 284: ...6 26 Operator s Manual 110017276EN Rev B 6 Troubleshooting ...
Page 329: ...Preface ...
Page 330: ...Table of Contents ...
Page 331: ...Getting Started ...
Page 332: ...User Interface ...
Page 333: ...Customizing Your System ...
Page 334: ...Detailed Reference ...
Page 335: ...Cleaning and Sterilization ...
Page 336: ...Troubleshooting ...
Page 337: ...Service and Warranty ...
Page 338: ...Specifications ...
Page 339: ...Glossary ...
Page 340: ...Index ...
Page 341: ...Preface ...
Page 342: ...Table of Contents ...
Page 343: ...Getting Started ...
Page 344: ...User Interface ...
Page 345: ...Customizing Your System ...
Page 346: ...Detailed Reference ...
Page 347: ...Cleaning and Sterilization ...
Page 348: ...Troubleshooting ...
Page 349: ...Service and Warranty ...
Page 350: ...Specifications ...
Page 351: ...Glossary ...
Page 352: ...Index ...