Operator’s Manual 6-5
110017276EN Rev. B
6 Troubleshooting
ID Type
Text Message
Suggested Action(s)
CPX06
Warning
The air pressure output is higher than
commanded.
SG46 Check the air or fluid infusion
pressure to be suitable for use in surgery.
SG04 Call your product service
representative if this problem persists.
CPX08
Warning
The internal air pressure system cannot
reach the full pressure expected.
SG25 Confirm correct infusion and
aspiration function.
SG31 Confirm correct vitrectomy function.
SG04 Call your product service
representative if this problem persists.
CPX09
Warning
The internal vitrectomy air pressure is
low.
SG31 Confirm correct vitrectomy function.
SG04 Call your product service
representative if this problem persists.
CPX10 Warning
Viscous fluid injection pressure is low.
SG91 Confirm correct viscous fluid
injection function.
SG04 Call your product service
representative if this problem persists.
CPX11 Warning
Viscous fluid injection pressure is high.
SG91 Confirm correct viscous fluid
injection function.
SG04 Call your product service
representative if this problem persists.
CPX12 Warning
Viscous fluid extraction vacuum is low.
SG92 Confirm correct viscous fluid
extraction function.
SG04 Call your product service
representative if this problem persists.
CPX13 Warning
Viscous fluid extraction vacuum is high.
SG92 Confirm correct viscous fluid
extraction function.
SG04 Call your product service
representative if this problem persists.
CPX14 Warning
Source pressure is low.
SG90 Increase input pressure to 72.5 to
100 psi. (5.0 bar - 6.9 bar)
CPX15
Warning
The internal air pressure supply to the
vacuum fluidics module is low. The
irrigation and aspiration pinch valves
may not be functioning properly.
SG25 Confirm correct infusion and
aspiration function.
SG04 Call your product service
representative if this problem persists.
CPX19
Warning
The vitrectomy cutter input pressure
is high. The vitrectomy cutter is not
available.
SG04 Call your product service
representative if this problem persists.
CPX20
Warning
The internal air pressure supply to the
vacuum fluidics module is high. The
irrigation and aspiration pinch valves
may not be functioning properly.
SG25 Confirm correct infusion and
aspiration function.
SG04 Call your product service
representative if this problem persists.
Summary of Contents for STELLARIS PC
Page 1: ...Operator s Manual ...
Page 78: ...1 68 Operator s Manual 110017276EN Rev B 1 Getting Started ...
Page 190: ...3 36 Operator s Manual 110017276EN Rev B 3 ...
Page 244: ...4 54 Operator s Manual 110017276EN Rev B 4 ...
Page 284: ...6 26 Operator s Manual 110017276EN Rev B 6 Troubleshooting ...
Page 329: ...Preface ...
Page 330: ...Table of Contents ...
Page 331: ...Getting Started ...
Page 332: ...User Interface ...
Page 333: ...Customizing Your System ...
Page 334: ...Detailed Reference ...
Page 335: ...Cleaning and Sterilization ...
Page 336: ...Troubleshooting ...
Page 337: ...Service and Warranty ...
Page 338: ...Specifications ...
Page 339: ...Glossary ...
Page 340: ...Index ...
Page 341: ...Preface ...
Page 342: ...Table of Contents ...
Page 343: ...Getting Started ...
Page 344: ...User Interface ...
Page 345: ...Customizing Your System ...
Page 346: ...Detailed Reference ...
Page 347: ...Cleaning and Sterilization ...
Page 348: ...Troubleshooting ...
Page 349: ...Service and Warranty ...
Page 350: ...Specifications ...
Page 351: ...Glossary ...
Page 352: ...Index ...