1-58 Operator’s Manual
110017276EN Rev. B
1 Getting Started
Linear Coagulation Control
The control power is varied linearly from preset minimum to the preset limit. Power begins when entering
footpedal position 1 and ends at the completion of travel.
Note:
Due to compliance with IEC 60601-2-2, position 1 will not start until approximately
35% of pedal travel is attained in the linear coagulation mode.