4-32 Operator’s Manual
110017276EN Rev. B
4 Detailed Reference
3.
Select
Surgeon’s Name
and select
Confirm
.
4.
Open the disposable package and insert the fluidics cassette all the way in and hold until it is
automatically captured by the system. The cassette housing backlight will stop blinking and remain on
when the system captures the cassette.
5.
After the vacuum test completes, the
Setup Screen
will display with
Prime and Tune
as the highlighted
function.
6.
Spike the BSS bottle and hang it at the desired bottle height.
Additional step if pressurized infusion is used:
Connect the Air Tubing Line to the vent port at the bottle spike and the other end with air filter to
the
Stellaris® PC
air output connector. Switch on the air pump from the system setup screen (use the
control in the upper right hand corner of the screen). The output connector will remain lit when it is at
commanded pressure, and blink on and off when it is not at the commanded pressure.
7.
Connect tubing to the I/A handpieces. Select
Show Me Steps
for animated setup guide, if necessary.
8.
Ensure the irrigation clamp is open and toggle
Irrigation Off
to turn the flow on and allow irrigation to
fill the tubing up to the handpieces.
9.
Fill the test chamber with irrigating solution, then slide over the tip of the handpiece.
10.
Select
Prime Only
. A vacuum test is part of the priming cycle.
11. After successful priming and tuning, the Main Surgical Screen will be displayed.
The external components of your system are now ready. Continue to set the operating parameters.
Note:
Hold the handpiece tip towards the ceiling while priming the irrigation line to insure all air has
been removed.
I/A Use
1.
Select the I/A mode on the clock menu.
2.
Use the spin control buttons to set the desired aspiration vacuum.
3.
The system is now ready for Irrigation/Aspiration.
Note:
See Chapter 5 for cleaning and sterilization requirements when surgery is completed.
Ultrasound Function
Phacoemulsification refers to the process of ultrasonic disintegration of the lens using a vibrating needle
operating at a frequency above the audible range, in the anterior chamber of the eye.
Summary of Contents for STELLARIS PC
Page 1: ...Operator s Manual ...
Page 78: ...1 68 Operator s Manual 110017276EN Rev B 1 Getting Started ...
Page 190: ...3 36 Operator s Manual 110017276EN Rev B 3 ...
Page 244: ...4 54 Operator s Manual 110017276EN Rev B 4 ...
Page 284: ...6 26 Operator s Manual 110017276EN Rev B 6 Troubleshooting ...
Page 329: ...Preface ...
Page 330: ...Table of Contents ...
Page 331: ...Getting Started ...
Page 332: ...User Interface ...
Page 333: ...Customizing Your System ...
Page 334: ...Detailed Reference ...
Page 335: ...Cleaning and Sterilization ...
Page 336: ...Troubleshooting ...
Page 337: ...Service and Warranty ...
Page 338: ...Specifications ...
Page 339: ...Glossary ...
Page 340: ...Index ...
Page 341: ...Preface ...
Page 342: ...Table of Contents ...
Page 343: ...Getting Started ...
Page 344: ...User Interface ...
Page 345: ...Customizing Your System ...
Page 346: ...Detailed Reference ...
Page 347: ...Cleaning and Sterilization ...
Page 348: ...Troubleshooting ...
Page 349: ...Service and Warranty ...
Page 350: ...Specifications ...
Page 351: ...Glossary ...
Page 352: ...Index ...