![Bausch & Lomb STELLARIS PC Operator'S Manual Download Page 209](http://html.mh-extra.com/html/bausch-and-lomb/stellaris-pc/stellaris-pc_operators-manual_2711462209.webp)
Operator’s Manual 4-19
110017276EN Rev. B
4 Detailed Reference
Note:
Make sure to dispose of the old lamp properly. Lamps are filled with high pressure xenon gas
or xenon and mercury gases. When disposing of the used lamp, take appropriate measures
in compliance with applicable regulations regarding waste disposal, or entrust disposal to a
licensed industrial waste disposal company. Be sure to comply with the regulations in your
country, state, region or province to ensure the used lamp is disposed of legally and correctly.
Fluid/Air Exchange (F/AX) (Posterior Only)
The Fluid/Air Exchange function provides continuous air flow from a pneumatic pump located in the
compressor. The air output connector must be used in the posterior mode for Fluid/Air Exchange. Air is filtered
through a disposable filter. The Foot Control side button may be programmed to turn Fluid/Air Exchange On or
Off.
Fluid/Air Exchange Operation
The Fluid/Air Exchange mode is a posterior feature which provides the surgeon with the ability to instill a
preset air pressure into the eye for Fluid/Air Exchange. It supports pressures up to 150 mmHg at flow rates
up to 3.5 standard cubic feet (99 liters) per hour. The air pressure from the pump may be adjusted from 0 to
150 mmHg in 1 mmHg increments.
Fluid/Air Exchange Setup and Use
The Fluid/Air Exchange tubing is an integrated part of the pre-connected tubing set of the
Stellaris® PC
Vision
Enhancement System procedure pack. The function shares the same air source as AFI. In the event the air pump
is already on for AFI, all you need to do to activate Fluid/Air Exchange is switch the 2-way stopcock that is
attached to the infusion cannula.
1.
Setup the system for posterior vitrectomy (see page 4-7 ) or combined surgery (see page 4-50 ).
2.
Select the On/Off button to turn On air pump. (Scrub nurse or Circulating nurse)
Note:
When the pump is turned on, the air pressure numeric display will transition to show actual
pressure. When the pump is off, the display indicates the preset air pressure.
Summary of Contents for STELLARIS PC
Page 1: ...Operator s Manual ...
Page 78: ...1 68 Operator s Manual 110017276EN Rev B 1 Getting Started ...
Page 190: ...3 36 Operator s Manual 110017276EN Rev B 3 ...
Page 244: ...4 54 Operator s Manual 110017276EN Rev B 4 ...
Page 284: ...6 26 Operator s Manual 110017276EN Rev B 6 Troubleshooting ...
Page 329: ...Preface ...
Page 330: ...Table of Contents ...
Page 331: ...Getting Started ...
Page 332: ...User Interface ...
Page 333: ...Customizing Your System ...
Page 334: ...Detailed Reference ...
Page 335: ...Cleaning and Sterilization ...
Page 336: ...Troubleshooting ...
Page 337: ...Service and Warranty ...
Page 338: ...Specifications ...
Page 339: ...Glossary ...
Page 340: ...Index ...
Page 341: ...Preface ...
Page 342: ...Table of Contents ...
Page 343: ...Getting Started ...
Page 344: ...User Interface ...
Page 345: ...Customizing Your System ...
Page 346: ...Detailed Reference ...
Page 347: ...Cleaning and Sterilization ...
Page 348: ...Troubleshooting ...
Page 349: ...Service and Warranty ...
Page 350: ...Specifications ...
Page 351: ...Glossary ...
Page 352: ...Index ...