4-30 Operator’s Manual
110017276EN Rev. B
4 Detailed Reference
Irrigation/Aspiration Function (I/A)
Note:
Do not pull the tubing taut — it must be allowed to have a droop or sag between the cassette
and the handpiece.
Note:
Whenever the cassette is ejected from the system console, keep the handpiece above the level of
the cassette port.
The Advanced Vacuum Function provides vacuum levels from 0 mmHg to 600 mmHg in 1 mmHg increments
depending upon the mode of operation. Aspiration limits are set via the touch screen, the remote control, or the
Foot Control buttons (if programmed).
In I/A mode, irrigation is activated in Region 1 of footpedal travel, and both irrigation and aspiration are
activated in Region 2 of footpedal travel.
Capsule Polish
The capsule polish function is typically accomplished with a lower vacuum setting than standard settings. These
settings may be customized to allow quick entry into a lower vacuum level as explained in Chapter 3 .
Viscoelastic Removal
The viscoelastic removal function provides different settings for the doctor’s convenience. These settings may
be customized to allow quick entry into a specific vacuum level as explained in Chapter 3 .
Venting
The Vacuum Flow Fluidics system provides the surgeon with either air or fluid venting options to free an
occluded tip when the footpedal is released. When air venting is selected, the residual vacuum is vented to
atmospheric pressure, and when fluid venting is selected, it is vented to a positive pressure equal to the bottle
height head pressure. Venting automatically occurs every time the Footpedal travels back to region 1.
Reflux
Aspiration of fluid to the collection cassette occurs via the handpiece and a tube set. Reflux applies a
momentary reverse pressure through the aspiration line to clear the aspiration port of lodged material.
The
Stellaris® PC
Vision Enhancement System is designed for use with continuous reflux or pulse reflux. If
enabled, the reflux feature is activated by inward movement of the footpedal in all aspiration modes.
Continuous Reflux:
Reflux pressure is based on irrigation pressure.
Summary of Contents for STELLARIS PC
Page 1: ...Operator s Manual ...
Page 78: ...1 68 Operator s Manual 110017276EN Rev B 1 Getting Started ...
Page 190: ...3 36 Operator s Manual 110017276EN Rev B 3 ...
Page 244: ...4 54 Operator s Manual 110017276EN Rev B 4 ...
Page 284: ...6 26 Operator s Manual 110017276EN Rev B 6 Troubleshooting ...
Page 329: ...Preface ...
Page 330: ...Table of Contents ...
Page 331: ...Getting Started ...
Page 332: ...User Interface ...
Page 333: ...Customizing Your System ...
Page 334: ...Detailed Reference ...
Page 335: ...Cleaning and Sterilization ...
Page 336: ...Troubleshooting ...
Page 337: ...Service and Warranty ...
Page 338: ...Specifications ...
Page 339: ...Glossary ...
Page 340: ...Index ...
Page 341: ...Preface ...
Page 342: ...Table of Contents ...
Page 343: ...Getting Started ...
Page 344: ...User Interface ...
Page 345: ...Customizing Your System ...
Page 346: ...Detailed Reference ...
Page 347: ...Cleaning and Sterilization ...
Page 348: ...Troubleshooting ...
Page 349: ...Service and Warranty ...
Page 350: ...Specifications ...
Page 351: ...Glossary ...
Page 352: ...Index ...