6 Troubleshooting
Vacuum Fluidics Module Messages
ID Type
Text Message
Suggested Action(s)
VFM01
Warning
The vacuum fluidics module was not
detected in the system. Surgical mode
with the vacuum fluidics module is not
available.
SG01 Call your product service
representative.
VFM02
Warning
The vacuum fluidics module software
version is not compatible with this
software version. Surgical mode is not
available.
SG02 Call your product service
representative. A compatible software
version must be downloaded.
VFM03
Warning
The vacuum fluidics module has failed
to respond to a settings command. The
module settings have been re-sent to the
module.
SG03 Confirm correct infusion, aspiration,
and vitrectomy function.
SG04 Call your product service
representative if this problem persists.
VFM04
Warning
The vacuum fluidics module has reset.
The module settings have been re-sent to
the module.
SG03 Confirm correct infusion, aspiration,
and vitrectomy function.
SG04 Call your product service
representative if this problem persists.
VFM05
Warning
The vacuum fluidics module does not
have valid factory calibration data.
Surgical mode is not available.
SG05 Call your product service
representative to replace the vacuum
fluidics modules.
SG06 Call your product service
representative to replace or calibrate the
vacuum fluidics module.
VFM07
Info
The vacuum fluidics module cassette
vacuum check is in progress.
SG13 Please wait for the vacuum fluidics
module cassette vacuum check to complete.
VFM08
Info
The cassette vacuum check has failed.
SG11 Re-insert the cassette for the cassette
vacuum check.
SG15 Replace the cassette.
SG04 Call your product service
representative if this problem persists.
VFM09
Info
System priming has not been completed.
SG20 Select ‘Setup’, and select ‘Prime’ to
complete the system priming.
VFM10
Info
System priming is in progress.
SG21 Please wait for the system priming to
complete.
VFM11
Info
System priming has failed.
SG22 Check irrigation and aspiration
tubing connections for leaks. Check that
the test chamber is forming a tight seal
around the handpiece. Try priming the
system again.
SG23 Eject the cassette, load a new pack,
and try priming the system again.
SG04 Call your product service
representative if this problem persists.
Summary of Contents for STELLARIS PC
Page 1: ...Operator s Manual ...
Page 78: ...1 68 Operator s Manual 110017276EN Rev B 1 Getting Started ...
Page 190: ...3 36 Operator s Manual 110017276EN Rev B 3 ...
Page 244: ...4 54 Operator s Manual 110017276EN Rev B 4 ...
Page 284: ...6 26 Operator s Manual 110017276EN Rev B 6 Troubleshooting ...
Page 329: ...Preface ...
Page 330: ...Table of Contents ...
Page 331: ...Getting Started ...
Page 332: ...User Interface ...
Page 333: ...Customizing Your System ...
Page 334: ...Detailed Reference ...
Page 335: ...Cleaning and Sterilization ...
Page 336: ...Troubleshooting ...
Page 337: ...Service and Warranty ...
Page 338: ...Specifications ...
Page 339: ...Glossary ...
Page 340: ...Index ...
Page 341: ...Preface ...
Page 342: ...Table of Contents ...
Page 343: ...Getting Started ...
Page 344: ...User Interface ...
Page 345: ...Customizing Your System ...
Page 346: ...Detailed Reference ...
Page 347: ...Cleaning and Sterilization ...
Page 348: ...Troubleshooting ...
Page 349: ...Service and Warranty ...
Page 350: ...Specifications ...
Page 351: ...Glossary ...
Page 352: ...Index ...