Operator’s Manual 6-21
110017276EN Rev. B
6 Troubleshooting
Symptom
Action Required
GUI hung up and
system ceased
operation
When the problem occurs, the system will automatically turn ON the irrigation valve
to provide continuous irrigation. System and MMC need to be restarted to resume
operation.
To restart system, perform the following:
A. Power down system by pressing and holding the standby power switch in front of
the system.
B. Wait until the system powers down completely.
C. Power down MMC by pressing and holding the power switch until the light turns
brighter (if not already powered down).
D. Restart MMC by pressing the power switch.
E. Restart system by pressing the standby power switch once.
F. Repeat system set up sequence,
Prime and Tune
system to resume operation.
6.5.
System Configurations, Modules, Accessories and
Packs
Use of non-approved accessories, packs or parts may affect system performance. The unauthorized modification
or alteration of the equipment, or the use of non-approved accessories, packs or parts with the equipment shall
relieve Lomb from any warranty, service obligation or other liability for damages to, or failure of, the
equipment caused by such unauthorized acts.
For additional information, please contact a Lomb sales representative or your local sales office.
The system configurations, modules, accessories and packs provided on the following pages are to be used
with the
Stellaris® PC
Vision Enhancement System . Although not provided in the list; the ultrasound needles,
I/A handpiece, I/A tips, and bipolar accessories for the
Millennium
Microsurgical System and Stellaris® are
compatible and could be used with the
Stellaris® PC
Vision Enhancement System .
Available Accessories and Packs
The packs listed below can be used with the
Stellaris® PC
Vision Enhancement System . A complete list of
accessories, including approved needles and sleeves and packs is available from your sales representative upon
request.
Summary of Contents for STELLARIS PC
Page 1: ...Operator s Manual ...
Page 78: ...1 68 Operator s Manual 110017276EN Rev B 1 Getting Started ...
Page 190: ...3 36 Operator s Manual 110017276EN Rev B 3 ...
Page 244: ...4 54 Operator s Manual 110017276EN Rev B 4 ...
Page 284: ...6 26 Operator s Manual 110017276EN Rev B 6 Troubleshooting ...
Page 329: ...Preface ...
Page 330: ...Table of Contents ...
Page 331: ...Getting Started ...
Page 332: ...User Interface ...
Page 333: ...Customizing Your System ...
Page 334: ...Detailed Reference ...
Page 335: ...Cleaning and Sterilization ...
Page 336: ...Troubleshooting ...
Page 337: ...Service and Warranty ...
Page 338: ...Specifications ...
Page 339: ...Glossary ...
Page 340: ...Index ...
Page 341: ...Preface ...
Page 342: ...Table of Contents ...
Page 343: ...Getting Started ...
Page 344: ...User Interface ...
Page 345: ...Customizing Your System ...
Page 346: ...Detailed Reference ...
Page 347: ...Cleaning and Sterilization ...
Page 348: ...Troubleshooting ...
Page 349: ...Service and Warranty ...
Page 350: ...Specifications ...
Page 351: ...Glossary ...
Page 352: ...Index ...