Asterion DC Multioutput Series
Programming Manual
13-126
M330517-01 Rev C
CAL<n>:UNLOCK “6867”
4.
Program the overvoltage protection to maximum to prevent nuisance trips:
CAL<n>:OUTP:VOLT:PROT:DAC 65535
5.
Program the output current to full scale to prevent Constant-Current operation:
CAL<n>:OUTP:CURR:DAC 65535
6.
Set the output voltage sensing to “Remote”
OUTP<n>:SENS REMOTE
7.
Set the output ON
OUTP<n>:STAT ON
8.
Program the output of the first calibration point by sending the following
command string from the computer:
CAL<n>:OUTP:VOLT:DAC 3275
9.
Let the output settle and measure the voltage with a high precision voltmeter;
this is value 1.
10.
Enter the actual output voltage value of the first calibration point:
CAL<n>:OUTP:VOLT:FIVEPOINT 1 <value 1>
11.
Program the output of the second calibration point by sending the following
command string from the computer:
CAL<n>:OUTP:VOLT:DAC 19000
12.
Let the output settle and measure the voltage with a high precision voltmeter;
this is value 2.
13.
Enter the actual output voltage value of the second calibration point:
CAL<n>:OUTP:VOLT:FIVEPOINT 2 <value 2>
14.
Program the output of the third calibration point by sending the following
command string from the computer:
Summary of Contents for Sorensen Asterion DC ASM
Page 1: ...M330517 01 Rev C Asterion DC Multioutput Series Programming Manual...
Page 6: ...Asterion DC Multioutput Series Programming Manual M330517 01 Rev C v SAFETY SYMBOLS...
Page 145: ...Asterion DC Multioutput Series Programming Manual 13 128 M330517 01 Rev C CAL n LOCK...
Page 162: ...M330461 01 Rev C 14 145...