Asterion DC Multioutput Series
Programming Manual
13-132
M330517-01 Rev C
4.
To calibrate channel 2, Connect zero volt (0 +/- 0.003V) across PIN 1
(VPRG2_VSOUR) and PIN 3 (PRG_RTN) of channel 2 Analog Programming
connector.
5.
To calibrate channel 3, Connect zero volt (0 +/- 0.003V) across PIN 1
(VPRG3_VSOUR) and PIN 3 (PRG_RTN) of channel 3 Analog Programming
connector.
6.
Select the Analog programming source as Voltage source and 10V range. Also
select programming mode as internal during calibration procedure.
SOUR<n>:ANALOG:PROG:SOUR VOLT
SOUR<n>:VOLT:PROG INT
SOUR<n>:VOLT:PROG:FSC 10
7.
Program the supply to unlock the non-volatile memory for calibration value
storage:
CAL<n>:UNLOCK “6867”
8.
Calibrate the Offset point.
CAL<n>:OUTP:VOLT:EXTV:OFFS 0
9.
To calibrate channel 1, Set the DC voltage to 10V +/-0.003V at PIN1
(VPRG1_VSOUR) of channel 1 Analog Programming connector. Calibrate the
full scale point.
10.
To calibrate channel 2, Set the DC voltage to 10V +/-0.003V at PIN1
(VPRG2_VSOUR) of channel 2 Analog Programming connector. Calibrate the
full scale point.
11.
To calibrate channel 3, Set the DC voltage to 10V +/-0.003V at PIN1
(VPRG3_VSOUR) of channel 3 Analog Programming connector. Calibrate the
full scale point.
CAL<n>:OUTP:VOLT:EXTV:FSC <100% channel rating voltage>
12.
Store the Calibration.
CAL<n>:STORE
CAL<n>:LOCK
13.6.2
Voltage programming
–
5V range voltage source
calibration
STEP
DESCRIPTION
1.
Issue a *CLS command.
2.
Issue a *RST command.
Summary of Contents for Sorensen Asterion DC ASM
Page 1: ...M330517 01 Rev C Asterion DC Multioutput Series Programming Manual...
Page 6: ...Asterion DC Multioutput Series Programming Manual M330517 01 Rev C v SAFETY SYMBOLS...
Page 145: ...Asterion DC Multioutput Series Programming Manual 13 128 M330517 01 Rev C CAL n LOCK...