Asterion DC Multioutput Series
Programming Manual
M330517-01 Rev C
4-11
Remote Programming Via RS232
4.2.1
RS232 Connector Pinout
The RS232 interface operates at a default baud of 9600. The baud is selectable from
9600 to 115200.
The RS232 interface is accessible through the rear panel DB9 connector (Refer to
), labeled Serial on the power supply’s rear panel (see
Figure 4-2: RS-232C Interface Connector
Connector
Type
RS-232C Interface
9-contact receptacle (female) Subminiature-D.
Table 4-1: RS-232C Interface Connector Type
Pin #
Name
DCE Signal
Direction
1
N/C
N/A
N/A
2
TxD
Transmit Data
Output
3
RxD
Receive Data
Input
4
N/C
N/A
N/A
5
Common
N/A
N/A
6
N/C
N/A
N/A
7
RTS
Request to Send
Input
8
CTS
Clear to Send
Output
9
N/C
N/A
N/A
Table 4-2: RS-232C Interface Connector Pinout
4.2.2
RS232 Setup Procedure
This section provides a quick reference for the configuration requirements for RS232.
Parameter
Setting
Notes
Baud Rate
Selectable from 9600 to 115200
The baud rate is
selectable through the
front panel
Data Bits
8
Not Selectable
Stop Bits
1
Not Selectable