Asterion DC Multioutput Series
Programming Manual
M330517-01 Rev C
10-89
Command
Description
SCPI
CALibrate<n>:INITial:MEASure:CURRent
:AVErage <NR1>
Sets the number of readings to
average together when returning the
current value with the
MEAS<n>:CURR? command to
reduce noise in the readback
readings. Enter a value of 3 to 9,
with the value of 3 (factory default)
providing the fastest response time
in the readings, but less rejection of
noise.
N
CALibrate<n>:INITial:MEASure:CURRent
:AVErage?
Returns the number 3 to 9 to
indicate the number of readings to
average together when taking a
current reading.
N
CALibrate<n>:INITial:MEASure:VOLTage
:AVErage <NR1>
Sets the number of readings to
average together when returning the
voltage value with the
MEAS<n>:VOLT? command to
reduce noise in the readback
readings. Enter a value of 1 to 10,
with the value of 1 (factory default)
providing the fastest response time
in the readings, but less rejection of
noise.
N
CALibrate<n>:INITial:MEASure:VOLTage
:AVErage?
Returns the number 1 to 10 to
indicate the number of readings to
average together when taking a
current reading.
N
CALibrate<n>:INITial:OUTPut:PROTecti
on:
FOLD <NR1>
Sets the power-on default foldback
protection setting. Valid arguments
are same as for
OUTP<n>:PROT:FOLD
N
CALibrate<n>:INITial:OUTPut:PROTecti
on:FOLD?
Returns the power-on default setting
of foldback protection
N
CALibrate<n>:INITial:REMote:INHIBIT
<0|1>
Sets the power-on default Remote
Inhibit Mode.
N
CALibrate<n>:INITial:REMote:INHIBIT?
Returns the power-on default
Remote Inhibit Mode settings.
N
CALibration<n>:INITial:ANALOG:REMote
:OUTPut <boolean>
Sets the power-on state of the
Remote output ON/OFF feature,
either enables or disables the
feature.
0 = Disables the Remote output
ON/OFF feature.
1 = Enables the Remote output
ON/OFF feature
N
CALibration<n>:INITial:ANALOG:REMote
:OUTPut?
Returns the default state of power-
on Remote ON/OFF feature
N
CALibrate<n>:MEASure:CURRent:CALCula
te
Calculates the value of the gain and
offset for current measurements.
N
CALibrate<n>:MEASure:CURRent:GAIN
<NRf>
Sets the value of the gain for current
measurements.
N
CALibrate<n>:MEASure:CURRent:GAIN?
Returns the value of the gain for
current measurements.
N
Summary of Contents for Sorensen Asterion DC ASM
Page 1: ...M330517 01 Rev C Asterion DC Multioutput Series Programming Manual...
Page 6: ...Asterion DC Multioutput Series Programming Manual M330517 01 Rev C v SAFETY SYMBOLS...
Page 145: ...Asterion DC Multioutput Series Programming Manual 13 128 M330517 01 Rev C CAL n LOCK...
Page 162: ...M330461 01 Rev C 14 145...