Error Code Description
-161
Invalid block data
The expected number of data values was not received.
-200
Execution error
An error/event number in the range [-299, -200] indicates that an error has
been detected by the instruments execution control block. The occurrence
of any error in this class shall cause the execution error bit (bit 4) in the
Event Status Register to be set. An execution error can be the result of:
•
A <program data> element out of range, such as programming 35 volts in
a 33-volt device.
•
A command could not be executed due to the current condition of the
device
.
-203
Command protected
Attempted to store calibration values to EEPROM without unlocking.
-221
Settings conflict
Attempted to set output greater than soft limits or to set soft limits less than
output
.
-222
Data out of range
Parameter exceeded range of valid values.
-241
Hardware missing
A legal command or query could not be executed because of a hardware
fault.
-340
Calibration failed
Error during calculation of calibration values occurred.
-350
Queue overflow
The error queue can contain up to 10 entries. If more than 10 error/event
conditions are logged before the SYSTem:ERRor? query, an overflow will
occur; the last queue entry will be overwritten with error -350. When the
queue overflows, the least recent error/events remain in the queue and the
most recent error/events are discarded.
-360
Communication error
Communications to a channel was disrupted.
Table 10-4: SCPI Error Codes
10.2.5
Serial Poll Operation
Performing a serial poll will not modify the SCPI Status Byte other than to clear the SRQ
(bit 6) for a Asterion DC Multioutput power Supply requesting service. Queries affecting
the Status Registers and subsequent serial poll are described below:
*ESR?
clears the ESR and bit 5 of the SCPI Status Register
SYSTem:ERRor?
clears bit 2 of the SCPI Status Register if the queue is empty
SCPI Conformance Information
The syntax of all SCPI commands implemented by the Asterion DC Multioutput Series
power supplies and documented in this manual, are either SCPI confirmed, or they are
customized commands not part of the SCPI definition. None of the commands
Summary of Contents for Sorensen Asterion DC ASM
Page 1: ...M330517 01 Rev C Asterion DC Multioutput Series Programming Manual...
Page 6: ...Asterion DC Multioutput Series Programming Manual M330517 01 Rev C v SAFETY SYMBOLS...
Page 145: ...Asterion DC Multioutput Series Programming Manual 13 128 M330517 01 Rev C CAL n LOCK...
Page 162: ...M330461 01 Rev C 14 145...