Asterion DC Multioutput Series
Programming Manual
12-112
M330517-01 Rev C
PROG
<n>
:DEF 7, VIMODE, 4, 5, 11, 6
//go to 4 volts, 5 amps, 11 volts
ovp, for 6 sec
PROG
<n>
:DEF 8, STOP
//stop running the sequence while the unit
remains at the state of the last command
within the sequence.
PROG
<n>
:STAT COMPLETE
PROG
<n>
:SAVE:SEL
PROG
<n>
:STAT?
Response:
State: STOPPED
OUTP
<n>
:STAT ON
PROG
<n>
:STAT RUN
The output of the supply for Example 3 should look like the following diagram:
Figure 12-3: Power Supply Output for Example 3
EXAMPLE 4
Issue
*RST
to establish a known state.
In this example, create a sequence that pauses at its end to let the last settings remain in
effect. To do this use the
PAUSE
command.
Now create a new SEQ3:
PROG
<n>
:
SEL:NAME “SEQ3”
PROG
<n>
:MALL DEFAULT
PROG
<n>
:DEF 1, VIMODE, 3, 4, 11, 10
//go to 3 volts, 4 amps, 11
volts ovp, for 10 sec
PROG
<n>
:DEF 2, RAMPTOV, 3, 5, 4, 11, 10
//ramp from 3 volts to 5
volts in 10 sec
3 volts for 10 sec
3 volts for 10 sec
5 volts for 10 sec
Ramp up to 5
volts in 10 sec
Ramp down to 3
volts in 10 sec
10 volts for 5 sec
Ramp down to 2
volts in 9 sec
4 volts for 6 sec
Summary of Contents for Sorensen Asterion DC ASM
Page 1: ...M330517 01 Rev C Asterion DC Multioutput Series Programming Manual...
Page 6: ...Asterion DC Multioutput Series Programming Manual M330517 01 Rev C v SAFETY SYMBOLS...
Page 145: ...Asterion DC Multioutput Series Programming Manual 13 128 M330517 01 Rev C CAL n LOCK...
Page 162: ...M330461 01 Rev C 14 145...