To calibrate channel 3, Connect 5k Ohm across PIN 5 (IPRG3_ISOUR) and
PIN 3 (PRG_RTN) of channel 3 Analog Programming connector.
CAL<n>:OUTP:CURR:EXTI:FSCR <100% channel rating current>
12.
Store the Calibration.
CAL<n>:STORE
CAL<n>:LOCK
REMOTE MONITOR OUTPUTS CALIBRATION
Remote Monitor output signals can be calibrated with set of commands as described in
below section. For detailed Pin description of the Remote Monitor Outputs and connection
diagram, Refer Section 8.4
13.7.1
Voltage Monitor Output Calibration
STEP
DESCRIPTION
1.
Issue a *CLS command.
2.
Issue a *RST command.
3.
Connect the DMM at PIN 7 (VMON1) and PIN 6 (MON_RTN) for channel 1,
PIN 7 (VMON2) and PIN 6 (MON_RTN) for channel 2 and PIN 7 (VMON3) and
PIN 6 (MON_RTN) for channel 3 to measure voltage monitor pin voltage and
Set the load for approximately 25% of rated output current respected channel.
Type the following commands.
SOUR<n>:VOLT 0
SOUR<n>:CURR <100% channel current rating>
OUTP<n>:STAT 1
CAL<n>
:UNLOCK “6867”
CAL<n>:INIT:VOLT:MON
4.
Calibrate the Offset point.
CAL<n>:OUTP:VOLT:MON:OFFS <voltage at PIN7 read from DMM>
5.
Set the output voltage to 80% of channel Full-scale rated voltage.
SOUR<n>:VOLT <0.8 times channel Full-scale voltage>
CAL<n>:OUTP:VOLT:MON:FSC <voltage at PIN7 read from DMM>
6.
Store the Calibration.
CAL<n>:STORE
CAL<n>:LOCK
Summary of Contents for Sorensen Asterion DC ASM
Page 1: ...M330517 01 Rev C Asterion DC Multioutput Series Programming Manual...
Page 6: ...Asterion DC Multioutput Series Programming Manual M330517 01 Rev C v SAFETY SYMBOLS...
Page 145: ...Asterion DC Multioutput Series Programming Manual 13 128 M330517 01 Rev C CAL n LOCK...
Page 162: ...M330461 01 Rev C 14 145...