Asterion DC Multioutput Series
Programming Manual
M330517-01 Rev C
13-125
CAL<n>:OUTP:VOLT:FIVEPOINT?
24. Send the command to calculate the calibration coefficients by the processor.
CAL<n>:OUTP:VOLT:CALC
25.
Program the supply to store the calibration values in non-volatile memory:
CAL<n>:STORE
Allow 15 seconds for the non-volatile memory to be updated with the
new calibration values.
26.
Program the supply to lock the non-volatile memory for calibration value
protection.
CAL<n>:LOCK
Voltage Programming Calibration
–
REMOTE
SENSE
Preparation:
1. Power down the Unit Under Test (UUT) and remove the input power for safety.
2. Connect output terminals to the Output connector on the rear side of the power
supply. Connect positive terminal to positive of Output connector and negative
terminal to negative of Output connector.
3. Connect the DVM to the UUT output terminals.
4. Apply input power and power up the UUT.
5. After the UUT has initialized, begin the calibration procedure.
STEP
DESCRIPTION
1.
Issue a *CLS command.
2.
Issue a *RST command.
3.
Program the supply to unlock the non-volatile memory for calibration value
storage: