Asterion DC Multioutput Series
Programming Manual
M330517-01 Rev C
13-123
Voltage Programming Calibration
STEP DESCRIPTION
1.
Issue a *CLS command.
2.
Issue a *RST command.
3.
Program the supply to unlock the non-volatile memory for calibration value
storage:
CAL<n>:UNLOCK “6867”
4.
Program the overvoltage protection to maximum to prevent nuisance trips:
CAL<n>:OUTP:VOLT:PROT:DAC 65535
5.
Program the output current to full scale to prevent Constant-Current operation:
CAL<n>:OUTP:CURR:DAC 65535
6.
Set the output voltage sensing to “Local”
OUTP<n>:SENS LOCAL
7.
Set the output ON
OUTP<n>:STAT ON
8.
Program the output of the first calibration point by sending the following
command string from the computer:
CAL<n>:OUTP:VOLT:DAC 3275
9.
Let the output settle and measure the voltage with a high precision voltmeter;
this is value 1.
10.
Enter the actual output voltage value of the first calibration point:
CAL<n>:OUTP:VOLT:FIVEPOINT 1 <value 1>
11.
Program the output of the second calibration point by sending the following
command string from the computer: