General Maintenance
4
Troubleshooting and Maintenance Manual
89
To refill/purge the EI calibration vial
Materials needed
•
PFTBA (05971-60571)
Refill
1
Stop any tuning or data acquisition.
2
Turn off the MS electronics.
3
Remove the analyzer window cover. See
4
Turn the calibration vial collar counterclockwise to loosen it (
).
Do not remove the collar.
5
Pull the calibration vial out. You may feel some resistance due to the
O-ring around the vial tube section.
6
Syringe or pipette PFTBA into the vial. With the vial vertical, the liquid
should be just below the end of the internal tube, approximately 70 µL of
sample.
7
Push the calibration vial into the valve as far as possible.
8
Withdraw the vial 1 mm. This prevents damage when you tighten the
collar.
Figure 7
Removing the EI calibration vial
Calibration vial
Calibration valve
Vent valve