General Maintenance
4
Troubleshooting and Maintenance Manual
109
To remove/reinstall the heater/sensor assembly of the ion source
Materials needed
•
Gloves, clean, lint-free
•
Large (8650-0030)
•
Small (8650-0029)
•
Hex ball driver, 1.5 mm (8710-1570)
•
Hex ball driver, 2.0 mm (8710-1804)
•
Hex nut driver, 5.5 mm (8710-1220)
Removal
1
Vent the MS. See the
Agilent 7000 Series Triple Quad GC/MS Operation
.
2
Open the front analyzer chamber. See the
Agilent 7000 Series Triple Quad
.
3
Remove the ion source. See “To remove the EI ion source” in the
7000 Series Triple Quad GC/MS Operation Manual
.
4
Remove the filaments.
WA R N I N G
The analyzer, GC/MS interface, and other components in the analyzer chamber
operate at very high temperatures. Do not touch any part until you are sure it is cool.
C A U T I O N
Always wear clean gloves to prevent contamination when working in the analyzer
chamber.
C A U T I O N
Make sure you use an antistatic wrist strap and take other antistatic precautions
before touching analyzer components.