![Agilent Technologies 7000 Triple Quadrupole GC/MS Maintenance Manual Download Page 41](http://html1.mh-extra.com/html/agilent-technologies/7000-triple-quadrupole-gc-ms/7000-triple-quadrupole-gc-ms_maintenance-manual_2867624041.webp)
General Troubleshooting
2
Troubleshooting and Maintenance Manual
41
•
MS electronics are not working correctly
* This will cause an error message.
GC/MS interface will not heat up
•
Incorrect setpoint(s)
•
Setpoint entered in wrong heated zone
•
GC/MS interface has not had enough time to reach temperature setpoint
•
GC is off
•
GC experienced a fault and needs to be reset*
•
GC/MS interface heater/sensor cable is not connected*
•
GC/MS heater failed (burned out)*
•
GC/MS sensor failed*
•
GC electronics are not working correctly*
* This will cause a GC error message. GC error messages are described in
the documentation supplied with your GC.