104
Troubleshooting and Maintenance Manual
4
General Maintenance
4
Use a clean, lint-free cloth or glove to wipe away excess grease. If the
O-ring looks shiny, there is too much grease on it.
5
Close the analyzer chamber.
6
Repeat this procedure for the other analyzer chamber.
7
Pump down the MS. See the
Agilent 7000 Series Triple Quad GC/MS
C A U T I O N
Do not use anything except the recommended vacuum grease. Excess grease can trap
air and dirt. Grease on surfaces of the O-ring other than the exposed surface can trap
air, resulting in air spikes during operation.
Figure 11
Side plate O-ring
Side plate O-rings