Troubleshooting and Maintenance Manual
4
General Maintenance
1
Vent the MS. See the
Agilent 7000 Series Triple Quad GC/MS Operation
.
2
Remove the column.
3
Disconnect the collision gas supply tubing.
4
Install a plug on the end of the collision gas tubing that leads into the
analyzer chamber.
5
Move the MS away from the GC.
6
Unplug the GC/MS interface heater cable from the GC.
7
Remove the front analyzer window and open the left side analyzer panel.
See
8
Finger-tighten the side plate thumbscrews for both analyzers. (
9
Switch the MS on to establish a rough vacuum.
10
When you hear the hissing sound of the pumpdown, close the vent valve.
Continue the pumpdown for 2 to 3 minutes.
11
Switch the MS off.
12
Close the analyzer cover and replace the front analyzer window.
13
Disconnect the LAN, remote, and power cable.
C A U T I O N
Do not overtighten the side plate thumbscrews. Overtightening will strip the threads in
the analyzer chamber. It will also warp the side plate and cause leaks.
C A U T I O N
Always wear clean gloves while handling any parts that go inside the analyzer
chambers.
Summary of Contents for 7000 Triple Quadrupole GC/MS
Page 24: ...24 Troubleshooting and Maintenance Manual 1 Introduction...
Page 50: ...50 Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Page 122: ...122 Troubleshooting and Maintenance Manual 4 General Maintenance...
Page 144: ...144 Troubleshooting and Maintenance Manual 6 Vacuum System...
Page 182: ...182 Troubleshooting and Maintenance Manual 8 Electronics...
Page 213: ......
Page 214: ...Agilent Technologies Agilent Technologies Inc Printed in USA September 2009...