General Troubleshooting
Lens supply fault
•
Electrical short in the analyzer
•
MS cannot maintain the voltage setpoint
•
MS electronics are not working correctly
Log amplifier ADC error
•
MS electronics are not working correctly
No peaks found
•
Emission current was set to 0
•
Electron multiplier voltage is too low
•
Poor mass axis calibration (either front or rear quad)
•
Width gain or offset is too high (either front or rear quad)
•
Calibration vial(s) empty or almost empty
•
Excessive pressure in the analyzer chambers
•
Air leak
•
Signal cable is not connected
•
Electrical leads to the detector are not connected correctly
•
HED power supply output cable failed
•
Electrical leads to the ion source are not connected correctly
•
Filament to the source body is shorted
Temperature control disabled
•
One of the heater fuses has failed
•
MS electronics are not working correctly
Temperature control fault
This indicates that something has gone wrong with the temperature control of
either the ion source or the mass filter (quad) heaters:
•
Source temperature sensor is open
•
Source temperature sensor is shorted
Summary of Contents for 7000 Triple Quadrupole GC/MS
Page 24: ...24 Troubleshooting and Maintenance Manual 1 Introduction...
Page 50: ...50 Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Page 122: ...122 Troubleshooting and Maintenance Manual 4 General Maintenance...
Page 144: ...144 Troubleshooting and Maintenance Manual 6 Vacuum System...
Page 182: ...182 Troubleshooting and Maintenance Manual 8 Electronics...
Page 213: ......
Page 214: ...Agilent Technologies Agilent Technologies Inc Printed in USA September 2009...