90
Troubleshooting and Maintenance Manual
4
General Maintenance
9
Turn the collar clockwise to tighten it.
The collar should be snug but not overly tight. Do
not
use a tool to tighten
the collar. It does not require that much force.
10
Reinstall the analyzer window cover.
11
In
Instrument Control
panel, select the
MS Tune
icon to display the
GC-QQQ
Tune
dialog box. Select the
Manual Tune
tab then select the
Ion Source
tab to
display the ion source parameters.
12
Turn off the
Emission
by clicking in the check box.
13
Purge the calibration valve by clicking in the
EI Cal Valve
check box to open
the calibration valve. Close the
EI Cal Valve
after 30 seconds.
C A U T I O N
After removing a calibrant vial, you
must
purge the calibration valve. Failure to do so
will result in damage to the filaments and the electron multiplier.
Summary of Contents for 7000 Triple Quadrupole GC/MS
Page 24: ...24 Troubleshooting and Maintenance Manual 1 Introduction...
Page 50: ...50 Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Page 122: ...122 Troubleshooting and Maintenance Manual 4 General Maintenance...
Page 144: ...144 Troubleshooting and Maintenance Manual 6 Vacuum System...
Page 182: ...182 Troubleshooting and Maintenance Manual 8 Electronics...
Page 213: ......
Page 214: ...Agilent Technologies Agilent Technologies Inc Printed in USA September 2009...