Introduction
1
Troubleshooting and Maintenance Manual
19
•
Turn off the hydrogen at its source every time you vent the MS (do not heat
the capillary column without carrier gas flow).
•
Turn off the hydrogen at its source every time shutoff valves in the MS are
closed (do not heat the capillary column without carrier gas flow).
•
Turn off the hydrogen at its source if a power failure occurs.
•
If a power failure occurs while the GC/MS system is unattended, even if the
system has restarted by itself:
1
Immediately turn off the hydrogen at its source.
2
Turn off the GC.
3
Turn off the MS and allow it to cool for 1 hour.
4
Eliminate
all
potential sources of ignition in the room.
5
Open the vacuum manifold of the MS to atmosphere.
6
Wait at least 10 minutes to allow any hydrogen to dissipate.
7
Start up the GC and MS as normal.
When using hydrogen gas
,
check the system for leaks to prevent possible fire
and explosion hazards based on local Environmental Health and Safety (EHS)
requirements. Always check for leaks after changing a tank or servicing the
gas lines. Always make sure the vent line is vented into a fume hood.
Summary of Contents for 7000 Triple Quadrupole GC/MS
Page 24: ...24 Troubleshooting and Maintenance Manual 1 Introduction...
Page 50: ...50 Troubleshooting and Maintenance Manual 2 General Troubleshooting...
Page 122: ...122 Troubleshooting and Maintenance Manual 4 General Maintenance...
Page 144: ...144 Troubleshooting and Maintenance Manual 6 Vacuum System...
Page 182: ...182 Troubleshooting and Maintenance Manual 8 Electronics...
Page 213: ......
Page 214: ...Agilent Technologies Agilent Technologies Inc Printed in USA September 2009...